Aero M dry powder dispersion system, designed to achieve the reproducible dispersion of bulk dry powder materials.
Includes a vibrating sample feed system, for sample delivery, along with a venturi disperser for sample dispersion. Manual selection of the dispersive air pressure (0-4bar) and sample feed rate is provided via software controls. Supplied, as standard, with a large volume sample feed tray (MAZ2063).
The Aero M unit automatically detects if the correct vacuum level is applied for sample extraction during a measurement, ensuring safe operation. Automatic control of a vacuum cleaner during a measurement sequence is not provided as standard, but requires purchase of a separate vacuum control unit (MAZ2080).
Flow path is constructed of 316 and hardened 410 stainless steel, borosilicate glass, EPDM, PTFE, Polyurethane, Aluminium, Carbon filled acetal, Neoprene.
Maximum measureable particle size is 1000 microns when used with the Mastersizer 3000E (sample dependant).
Requires, but does not include, a venturi disperser (MAZ2074-2075), a dry measurement cell (MAZ2050), a clean dry compressed air supply of 90psi (6 bar) at a flow rate of at least 90l/min, as provided by the COM0035/36 compressors and an industrial vacuum system, such as VAC0067/68, for sample collection.
NOTE:
(1) Any alternative to the VAC0067/68 must not be rated at more than 2.7kW @220v or 1.3kW @110v.
(2) In the event that any alternative to the COM0035/36 compressor is installed, it will be essential to install filter FIL0027 or equivalent in the airline. Failure to observe this requirement will invalidate the warranty for this dispersion unit.